Switch Motion Platforms
This page will list the steps required to switch from one motion platform to another when
only one 3030 scanhead is being used.
- Move the 3030 scanhead from the "old" motion platform to the "new" motion platform.
- Connect all cables from the "new" motion platform to the 3030 scan head.
- Verify that the file, 'deviceSetup.txt', is located in the
'...\Cyberware\CyScanÏgu\' folder. If it is not, please copy it from the
Cyberware Software CD-ROM and place it into this folder.
- Verify that all of your configuration files (for BOTH motion platforms) exist in the
'...\Cyberware\CyScanÏgu\' folder. If they are not present, please copy them from the
Cyberware Software CD-ROM and place them into this folder.
- Launch CyScan.
- Choose 'Scanner | Select Device...'.
![[Scanner Menu]](images/cyscan-004.png) |
| Scanner Menu |
- Click one time on the scanner motion platform that you intend to use.
- Click on the 'Properties...' button. You should see a window that looks very
similar to the image below. Note the buttons at the bottom of the window: 'Configure PSU' and/or 'Configure MSU'.
![[Choose a Scanner Dialog]](images/usb_scanhead_swap.jpg) |
| Choose a Scanner Dialog - Multiple Motion Platforms |
- Select the appropriate button (Configure PSU or Configuer MSU),
then wait for approximately 5 seconds while the information is uploaded to the scanhead EeProm chips.
- Exit CyScan.
- Restart CyScan to use the new settings.